At a glance
- Choose the agitator and incubator together: they work as one storage system.
- AP-48 L: up to 48 platelet bags or 16 apheresis bags. AP-96 L: up to 96 or 32.
- The LT incubators keep a programmable temperature between 18 °C and 22 °C with forced-air circulation.
- Add documentation to match your quality system: chart recorder, door-open and power-failure alarms, remote alarms and data logger.
Platelet storage needs continuous, uniform agitation and a stable temperature. Size both units for your expected inventory and operating procedures.
Key figures
- 48 / 96 platelet bags per agitator
- 16 / 32 apheresis bags
- 18–22 °C programmable in the LT incubators
1. Define your maximum inventory
Start from the maximum number of bags stored at the same time, not the daily average. Then add room for production peaks, turnover and growth: a system running close to its limit makes distribution harder on high-demand days.
Table: The AP range
- Model | Type | Capacity | Pairs with
- AP-48 L | Linear agitator | 48 platelet / 16 apheresis bags | AP-48 LT
- AP-96 L | Linear agitator | 96 platelet / 32 apheresis bags | AP-96 LT
- AP-48 LT | Incubator, 18–22 °C | Houses the AP-48 L | AP-48 L
- AP-96 LT | Incubator, 18–22 °C | Houses the AP-96 L | AP-96 L
2. Choose the agitator for your workflow
AP agitators move every unit with smooth, continuous linear motion on rollers, over perforated trays that let air circulate around the bags. The motor is low-maintenance with forced-air cooling, built for continuous operation.
Presvac tip
In high-demand services, one AP-96 L manages more units in a single device and avoids splitting inventory across several agitators.
3. Size the matching incubator
The incubator must house the agitator without blocking air circulation. The AP-48 LT pairs with the AP-48 L and the AP-96 LT with the AP-96 L. Both offer:
- Programmable temperature between 18 °C and 22 °C
- Digital temperature display
- Forced-air circulation for uniform temperature
- Visual and audible high and low temperature alarms
- Tempered glass door to check the contents without opening
- Stainless steel interior
4. Review alarms and monitoring
Table: Options by unit
- Unit | Optional features
- Agitators (AP-48 L / AP-96 L) | Stainless steel trays, motion alarm
- Incubators (AP-48 LT / AP-96 LT) | Chart recorder, door-ajar alarm, power-failure alarm, remote alarms, data logger
Choose options according to your traceability requirements and whether you need remote supervision or documented records.
5. Confirm installation and daily operation
Check the available space, access for loading and cleaning, the electrical supply and ambient conditions, and make sure the configuration holds the expected inventory without crowding.
Checklist before you buy
Checklist
- Maximum number of platelet and apheresis bags stored at once
- Expected growth over the next years
- Agitator and incubator sized as a pair
- Alarms and records your quality system requires
- Space, access and electrical supply in the storage area
Which configuration fits?
Which one fits
- AP-48 L + AP-48 LT Compact solution for lower volumes: up to 48 platelet bags or 16 apheresis bags.
- AP-96 L + AP-96 LT For high-demand blood banks: up to 96 platelet bags or 32 apheresis bags.
Frequently asked questions
FAQ
What temperature do the incubators keep?
They are programmable between 18 °C and 22 °C, with a digital display and high and low temperature alarms.
Is there an alarm if agitation stops?
Yes. A motion alarm is available as an option on the agitators.
Do you provide installation and training?
Yes. Presvac installs the equipment, trains the users and helps plan preventive and corrective maintenance.